SAR Efficient Imaging for Patients with Deep Brain Electrodes at 3 Tesla: Coil Length Matters
Motivation: To improve the design of pTx head coils intended to produce uniform RF fields while controlling local SAR levels for patients with DBS implants. Goal(s): To evaluate the effect of extended coil length and increased number of rows on performance using simulations. Approach: Comparison of SAR and RF homogeneity in 3-row and 4-row pTx head coil configurations using FDTD simulations on a human head model with an implanted DBS lead. Results: The 4-row coil offered the best SAR control for the whole brain, with length having a greater impact on local SAR than the number of rows. Impact: The research indicates that the length of multi-row pTx coils affects local SAR in MRI for DBS patients, enhancing efficiency and safety, and broadening imaging possibilities, potentially advancing MRI-compatible devices and personalized imaging techniques.
Authors
- Özlem İpek (ORCID: https://orcid.org/0000-0001-5233-1600)
- Joseph V. Hajnal (ORCID: https://orcid.org/0000-0002-2690-5495)
- Nejat Karadeniz (ORCID: https://orcid.org/0009-0003-9975-4469)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3728
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00