Mitigation of radiated E-field from a 3T MRI system operated without a Faraday shielded room using parallel transmission
Motivation: Eliminating the Faraday cage would lower installation costs by ~2X and facilitate deployment of MRI in diverse settings, but requires reducing the Tx system’s electromagnetic (EM) radiation. Goal(s): We employ a parallel transmit (pTx) array and EM absorbers to reduce RF-radiation from a 3T MRI operating without a shielded room, ensuring operation within regulatory limits. Approach: We model a 3T MRI with a CP body coil and design a 16-channel pTx array and EM-absorber to reduce E-field radiation. Performance is assessed with pTx pulse optimization and L-curve analysis. Results: We demonstrate a 2270-fold reduction in radiation compared to the birdcage coil. Impact: By successfully mitigating RF-radiation from MRI systems operated without Faraday cages, this research advances cost-effective MRI installations in diverse clinical/research environments, encouraging further exploration of pTx technology for controlling electromagnetic fields both inside/outside of the body.
Authors
- Ehsan Kazemivalipour (ORCID: https://orcid.org/0000-0003-4221-2397)
- Lawrence Wald
- Bastien Guérin (ORCID: https://orcid.org/0000-0002-4746-0886)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4087
- Citations
- 3
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 1.21