Denoise of Dynamic Sodium MRI imaging at 3T using MPPCA
Motivation: Sodium MRI, a valuable imaging modality for various clinical applications, faces challenges related to signal-to-noise ratio (SNR). This study aims to enhance sodium MRI image quality through innovative denoising techniques. Goal(s): Evaluate and compare novel denoising methods to improve sodium MRI quality and diagnostic potential. Approach: The study assesses denoising techniques in sodium MRI data, utilizing a 3T clinical scanner and a unique denoising algorithm MPPCA with reduced input requirements. Evaluation metrics include residual distributions, SNR and mean-square-error comparisons. Results: MPPCA consistently outperforms other methods, showing its effectiveness in reducing noise and maintaining data fidelity. Impact: The study demonstrates the potential of MPPCA as a powerful tool for enhancing sodium imaging, with broader implications for healthcare diagnostics and research.
Authors
- Ying‐Chia Lin (ORCID: https://orcid.org/0000-0001-7243-0980)
- Yongxian Qian (ORCID: https://orcid.org/0000-0002-9931-7790)
- Abhipsha Das
- Jelle Veraart (ORCID: https://orcid.org/0000-0003-0781-0420)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4206
- Primary Topic
- Nuclear Physics and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00