An Optimized Multi-Component Imaging Method on a Homebuilt 0.5 T MRI System: Combing Intra- and Inter-Voxel Constraints
Motivation: Applying the component imaging method to low-field MRI systems will face a main problem: low-SNR image data. Goal(s): An optimized inversion method is proposed, aiming to give better results for image data with low SNR. Approach: This paper proposes an optimized inversion method with the formula of the optimization problem combining intra- and inter-voxel constraints. Results: The optimized method shows a better convergence rate avoiding the fragmentation of component images and the appearance of pseudo peaks in the spectrum. Impact: This multi-component imaging approach can provide sub-voxel characterization and be applied to numerous applications of popular portable low-field MRI systems.
Authors
- Zhengxiu Wu (ORCID: https://orcid.org/0000-0002-3324-8534)
- Xiaowen Jiang (ORCID: https://orcid.org/0000-0002-3588-0357)
- Rongsheng Lu (ORCID: https://orcid.org/0000-0002-5802-7236)
- Zhonghua Ni (ORCID: https://orcid.org/0000-0001-9906-4546)
- Yi Chen (ORCID: https://orcid.org/0000-0002-5353-0609)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2845
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00