A MULTIPLEX-Lite method for fast 3D multi-parametric imaging with MR Angiography
Motivation: Existing 3D multi-parametric imaging methods suffers from long scan times, and decent MRA contrasts are yet to be provided. Goal(s): To develop a 3-minute 3D high resolution multi-parametric imaging method that offeres whole brain imaging with TOF quality MRA results. Approach: A simplified version of the MULTIPLEX method, namely MULTIPLEX-Lite, is proposed. Compared to the original MULTIPLEX method, addtional MRA results are offered, while removing the T1 and PD mappings to further improve scan time. Results: The proposed MULTIPLEX-Lite method offered whole brain high resolution (<1mm3 voxel volume) MRA, PDW, T1W, aT1W, SWI, R2* maps and QSM in one 3-minute scan. Impact: The proposed MULTIPLEX-Lite method is among the most feasible and practical solutions for routine clinically friendly 3D high-resolution multi-parametric imaging practices.
Authors
- Yongquan Ye (ORCID: https://orcid.org/0000-0002-6673-6753)
- Miaowen Li
- Jian Xu (ORCID: https://orcid.org/0000-0002-9370-4829)
- Hongyu Li (ORCID: https://orcid.org/0000-0001-9212-2010)
- Ying Wu (ORCID: https://orcid.org/0009-0006-0584-460X)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0570
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00