Safety of Intraoral Coils: Accuracy of Simulations
Motivation: In dental applications intraoral coils (IOCs) offer higher SNR than external surface coils, but the details of RF-induced heating of an intraoral coil dependent on many factors such as coil size, shape, incident E-field, surrounding tissue, and acquisition protocol. Goal(s): This study investigates the effect of anatomical model complexity on SAR simulations for IOCs. Approach: Numerical simulations were compared to electric field mapping and temperature measurements to evaluate RF-induced heating of various IOCs. Results: RF heating and SAR simulations of IOCs can be performed with limited tissue models, but the required complexity of the tissue model depends on the coil type. Impact: The complexity of numerical simulations can be reduced to limited tissues while maintaining accuracy of SAR estimation to determine the safety of intraoral coils through hotspot detection.
Authors
- Simon Reiss (ORCID: https://orcid.org/0000-0003-2449-1679)
- Michael Bock (ORCID: https://orcid.org/0000-0001-9720-3506)
- Agazi Tesfai (ORCID: https://orcid.org/0000-0002-6265-9682)
- Ali Çağlar Özen (ORCID: https://orcid.org/0000-0003-3536-0826)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3725
- Primary Topic
- Educational Robotics and Engineering
- Type
- article
- Field-Weighted Citation Impact
- 0.00