7 Tesla 31P Birdcage / 16-channel 1H Loop and Dipole Array integrated with Parallel Transmit System
Motivation: Integrating multinuclear applications into parallel transmit system (PTx) at ultra-high field MRI systems will provide uniform transmit field and good SNR for both proton and the x-nuclei. Goal(s): Design, build and integrate a 7T 31P/1H head array into a 7T PTx system that supports multinuclear applications. Approach: Use 8 loops and 8 folded dipoles as 1H transceivers, and a birdcage as 31P transceiver, for 31P/1H MRS and MRI in a 7T PTx systemResults: Phantom and in vivo experimental results show that the coil can provide homogeneous and good SNR signals for both 31P and 1H in ROIImpact: Elevating multinuclear studies with 7T parallel transmit system, providing uniform transmit fields and good receive sensitivity profile for both proton and x-nuclei in ROI. A potential game-changer for metabolic studies in body applications at ultra-high fields, advancing medical research.
Authors
- Ivan Dimitrov (ORCID: https://orcid.org/0000-0003-2686-6884)
- A Henning (ORCID: https://orcid.org/0000-0002-2267-4861)
- Wim Prins
- Paul Sanders (ORCID: https://orcid.org/0000-0003-3436-5334)
- Beichen Zhang (ORCID: https://orcid.org/0000-0002-0927-9568)
- Daniel Lowrance
- P. van der Meulen
- B. van den Berg
- Tom Geraedts
- Conrad Gohl
- Geert-Jan Plattel
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0463
- Primary Topic
- Wireless Body Area Networks
- Type
- article
- Field-Weighted Citation Impact
- 0.00