Real-time Respiratory Correction of the Pilot Tone Cardiac Signal to Stabilize Triggering during Breath-holds
Motivation: This research aims to refine Pilot Tone (PT) cardiac triggering, which can be impacted by changes in signal amplitude caused by breath-holding. Goal(s): Its primary goal is to develop and validate a real-time correction algorithm that improves the PT cardiac trigger stability during breath-holds without introducing additional variability during free breathing. Approach: The algorithm analyzes PT data for respiratory patterns to adapt beat-by-beat amplitude correction of the cardiac signal. The methodology was tested with different breathing paradigms at 1.5T and 3T to assess robustness. Results: The correction successfully adjusted PT cardiac signal amplitude, reducing missed triggers across respiratory patterns without increasing triggering variation. Impact: The described real-time correction improves the reliability of cardiac PT triggers during inspiratory breath-hold maneuvers, increasing the opportunities to replace ECG triggering with PT triggering, and therefore to eliminate the need for ECG leads.
Authors
- Orlando P. Simonetti (ORCID: https://orcid.org/0000-0002-8994-0095)
- Peter Speier
- Mario Bacher (ORCID: https://orcid.org/0000-0002-5529-3598)
- R. Badlishah Ahmad (ORCID: https://orcid.org/0000-0002-4862-2728)
- Yue Pan (ORCID: https://orcid.org/0009-0006-1421-9005)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4642
- Primary Topic
- Cardiovascular Syncope and Autonomic Disorders
- Type
- article
- Field-Weighted Citation Impact
- 0.00