New Dual-Element 13C/1H Endorectal Coil for Improving mpMR-TRUS Guided Fusion Prostate Biopsies With Hyperpolarized C-13 Molecular Imaging

Motivation: To address the issues with the original dual-element 13C/1H endorectal coil (ERC). Goal(s): We designed and fabricated a novel dual-element 13C/1H ERC, validated with bench testing and then evaluated in 3T MRI studies. Approach: Phantoms were imaged using the new ERC and the original ERC. Acquired data was then analyzed and respective SNR compared. Results: The new ERC demonstrated substantially improved 13C and 1H MRI sensitivity on bench electronic testing. In addition, the new ERC demonstrated substantial improvement in signal-to-noise ratio, and since the coil has successfully been used in hyperpolarized 13C-pyruvate, co-polarized 13C-urea+pyruvate multiparametric MRI patient studies for guiding prostate cancer biopsy. Impact: Hyperpolarized 13C MRI, RF Coils, Metabolic imaging, Molecular imaging, Endorectal coil

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4751
Primary Topic
Enzyme Structure and Function
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

New Dual-Element 13C/1H Endorectal Coil for Improving mpMR-TRUS Guided Fusion Prostate Biopsies With Hyperpolarized C-13 Molecular Imaging

Kimberly Okamoto, Mary Frost, Katsuto Shinohara, Daniel B. Vigneron et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Enzyme Structure and Function
article

New Dual-Element 13C/1H Endorectal Coil for Improving mpMR-TRUS Guided Fusion Prostate Biopsies With Hyperpolarized C-13 Molecular Imaging

Kimberly Okamoto, Mary Frost, Katsuto Shinohara, Daniel B. Vigneron, Peder E. Z. Larson, Yaewon Kim, Robert Bok, Jeremy W. Gordon, Michael A. Ohliger, Daniel Gebrezgiabhier, Rahul Aggarwal, Matthew R. Cooperberg, Zhen Wang, Hsin-Yu Chen, Hao Nguyen, Lucas Carvajal
article en

Abstract

Motivation: To address the issues with the original dual-element 13C/1H endorectal coil (ERC). Goal(s): We designed and fabricated a novel dual-element 13C/1H ERC, validated with bench testing and then evaluated in 3T MRI studies. Approach: Phantoms were imaged using the new ERC and the original ERC. Acquired data was then analyzed and respective SNR compared. Results: The new ERC demonstrated substantially improved 13C and 1H MRI sensitivity on bench electronic testing. In addition, the new ERC demonstrated substantial improvement in signal-to-noise ratio, and since the coil has successfully been used in hyperpolarized 13C-pyruvate, co-polarized 13C-urea+pyruvate multiparametric MRI patient studies for guiding prostate cancer biopsy. Impact: Hyperpolarized 13C MRI, RF Coils, Metabolic imaging, Molecular imaging, Endorectal coil

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Good health and well-being
Openalex Percentile: Top 87%
Enzyme Structure and Function
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies