Reduction of the RF-Induced Heating for the Partially-In and Partially-Out Medical Electrodes at 1.5T MRI

Motivation: RF-induced heating for the partially-in and partially-out (PIPO) medical implants is the primary concern for patient’s safety under MRI. Goal(s): Winding the external portion of the PIPO medical electrodes is proposed to reduce the RF heating under 1.5T MRI. Approach: One commercial PIPO medical electrode and an insulated solid wire were used to demonstrate this concept. RF heating is measured inside the ASTM phantom under a 1.5T RF birdcage coil and evaluated from the in-vivo simulations using the transfer function (TF) approach. Results: The results showed a significant reduction of RF heating in the commercial lead and simplified wire. Impact: The proposed method can reduce RF-induced heating significantly, mitigating the risk of tissue damage and improving the quality of care for patients with partially-in and partially-out (PIPO) medical devices under 1.5T MRI.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/3722
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reduction of the RF-Induced Heating for the Partially-In and Partially-Out Medical Electrodes at 1.5T MRI

Hongbae Jeong, Ran Guo, Jianfeng Zheng, Mir Khadiza Akter et al.
2 links
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Reduction of the RF-Induced Heating for the Partially-In and Partially-Out Medical Electrodes at 1.5T MRI

Hongbae Jeong, Ran Guo, Jianfeng Zheng, Mir Khadiza Akter, Md Zahidul Islam, Ananda Kumar, Ji Chen
article en

Abstract

Motivation: RF-induced heating for the partially-in and partially-out (PIPO) medical implants is the primary concern for patient’s safety under MRI. Goal(s): Winding the external portion of the PIPO medical electrodes is proposed to reduce the RF heating under 1.5T MRI. Approach: One commercial PIPO medical electrode and an insulated solid wire were used to demonstrate this concept. RF heating is measured inside the ASTM phantom under a 1.5T RF birdcage coil and evaluated from the in-vivo simulations using the transfer function (TF) approach. Results: The results showed a significant reduction of RF heating in the commercial lead and simplified wire. Impact: The proposed method can reduce RF-induced heating significantly, mitigating the risk of tissue damage and improving the quality of care for patients with partially-in and partially-out (PIPO) medical devices under 1.5T MRI.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.