3D-TSE Dixon sequence with preserved short echo spacing for second and subsequent echoes

Motivation: Common to existing TSE Dixon sequences is that the chemical shift between water and fat is encoded by shifting the readout window away from the spin-echo time point. The shifting prolongs the echo spacing by twice the shift. A short echo spacing is an important component of modern 3D-TSE sequences to enable high resolution 3D imaging in clinically acceptable times. Goal(s): Develop a 3D-TSE Dixon sequence with unchanged echo spacing. Approach: Encode the chemical shift via the excitation. Acquire in-phase and opposed-phase data for 2pt-Dixon symmetric around the spin-echo time point. Results: Homogeneous fat-water separation in knee, hand and ankle. Impact: 3D-TSE 2pt-Dixon sequence with unchanged echo spacing between second and subsequent echoes compared to a conventional 3D-TSE sequence with same acquisition parameters.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4567
Primary Topic
Seismic Imaging and Inversion Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

3D-TSE Dixon sequence with preserved short echo spacing for second and subsequent echoes

Alto Stemmer, Dominik M. Nickel, Andreas Schäfer
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Seismic Imaging and Inversion Techniques
article

3D-TSE Dixon sequence with preserved short echo spacing for second and subsequent echoes

Alto Stemmer, Dominik M. Nickel, Andreas Schäfer
article en

Abstract

Motivation: Common to existing TSE Dixon sequences is that the chemical shift between water and fat is encoded by shifting the readout window away from the spin-echo time point. The shifting prolongs the echo spacing by twice the shift. A short echo spacing is an important component of modern 3D-TSE sequences to enable high resolution 3D imaging in clinically acceptable times. Goal(s): Develop a 3D-TSE Dixon sequence with unchanged echo spacing. Approach: Encode the chemical shift via the excitation. Acquire in-phase and opposed-phase data for 2pt-Dixon symmetric around the spin-echo time point. Results: Homogeneous fat-water separation in knee, hand and ankle. Impact: 3D-TSE 2pt-Dixon sequence with unchanged echo spacing between second and subsequent echoes compared to a conventional 3D-TSE sequence with same acquisition parameters.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Clean water and sanitation
Openalex Percentile: Top 78%
Seismic Imaging and Inversion Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

3D-TSE Dixon sequence with preserved short echo spacing for second and subsequent echoes — Alto Stemmer, Dominik M. Nickel, et al. · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2024) | TGRS Research Map | TGRS