Efficient native MRA at 0.55T using motion-correction
Motivation: MRI at 0.55T with an 80cm bore has allowed improved access to CMR for patients. The application of robust native whole-thorax MRA at low-field remains limited especially due to the challenge of respiratory motion. Goal(s): To develop a large field-of-view, whole-thorax, self-gated and breathing-motion corrected native MRA with a predictable scan time and efficient data usage at 0.55T. Approach: An optimized sampling pattern is used with T2-prep, dual-echo Dixon and non-rigid motion correction is developed. Results: The approach was successfully applied in-vivo and shows successful mitigation of motion artefacts and blurring and good separation of water and fat in the entire thoracic region. Impact: Proof of the viability of self-gated MOCO MRA at 0.55T.
Authors
- Daniel Giese (ORCID: https://orcid.org/0000-0002-3925-4441)
- Christoph Kolbitsch
- Dominik Nickel (ORCID: https://orcid.org/0000-0002-0360-7233)
- Robert D. Stoll
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3365
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00