Enabling Wide-Area Imaging in MR Microscopy with a Double Helix Dipole Coil
Motivation: The DHD coil in the previous study had a narrow sensitivity region. We developed a DHD coil with improved sensitivity region and B1 uniformity to image large samples while maintaining SNR. Goal(s): Imaging of all organs, especially the brain of a human embryo, using a coil with a wider sensitivity region. Approach: We performed electromagnetic field simulations to find the shape that would give the best performance, and fabricated the coil. Results: The imageable area was widened by 30% and the insensitive area was reduced. Not only the brain of a human embryo, but also organs could be imaged at the same time. Impact: The DHD coil's geometry was determined by electromagnetic field simulation, which allows a 30% wider range in the z-axis than conventional coils. The human embryonic brain and organs such as the heart could be simultaneously imaged with sufficient SNR.
Authors
- Kazuki Kunieda
- Yasuhiko Terada (ORCID: https://orcid.org/0000-0002-8583-3057)
- Yuto Murakami
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1571
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00