Motion-induced B1+-changes in dynamic glucose enhanced (DGE) MRI and how to remedy them.

Motivation: Motion-induced B1+-changes at the voxel level result in erroneous dynamic glucose enhanced (DGE) MRI effects. Goal(s): To investigate effects of motion-induced B1+-changes on DGE MRI and to address removing them. Approach: A volunteer changed head positions, and voxel-based B1+ was measured pre- and post-motion. Z-spectra with and without D-glucose infusion were simulated, with and without the measured B1+-changes. Results: Slight motion-induced B1+-alterations lead to pseudo-CEST effects comparable to DGE effects. These can be removed by acquiring a full Z-spectrum and using the asymmetry of the glucoCEST signal relative to the water frequency to assess the DGE signal changes. Impact: Motion-induced B1+-changes affect DGE signals, thus causing pseudo-CEST effects that complicate clinical interpretation. These effects can be overcome by acquiring a full Z-spectrum and exploiting the asymmetry of the glucoCEST signal changes relative to the water frequency.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4481
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Motion-induced B1+-changes in dynamic glucose enhanced (DGE) MRI and how to remedy them.

Linda Knutsson, Nirbhay N. Yadav, Peter C.M. van Zijl, Karin Markenroth Bloch et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Motion-induced B1+-changes in dynamic glucose enhanced (DGE) MRI and how to remedy them.

Linda Knutsson, Nirbhay N. Yadav, Peter C.M. van Zijl, Karin Markenroth Bloch, Patrick Lehmann, Emil Ljungberg, Pia C. Sundgren, Ronnie Wirestam
article en

Abstract

Motivation: Motion-induced B1+-changes at the voxel level result in erroneous dynamic glucose enhanced (DGE) MRI effects. Goal(s): To investigate effects of motion-induced B1+-changes on DGE MRI and to address removing them. Approach: A volunteer changed head positions, and voxel-based B1+ was measured pre- and post-motion. Z-spectra with and without D-glucose infusion were simulated, with and without the measured B1+-changes. Results: Slight motion-induced B1+-alterations lead to pseudo-CEST effects comparable to DGE effects. These can be removed by acquiring a full Z-spectrum and using the asymmetry of the glucoCEST signal relative to the water frequency to assess the DGE signal changes. Impact: Motion-induced B1+-changes affect DGE signals, thus causing pseudo-CEST effects that complicate clinical interpretation. These effects can be overcome by acquiring a full Z-spectrum and exploiting the asymmetry of the glucoCEST signal changes relative to the water frequency.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Clean water and sanitation
Openalex Percentile: Top 63%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.