Improving Cardiac Phase-Resolved T2*-Mapping of the Murine Heart: Artifact Reduction and Enhanced Accuracy
Motivation: Rapid blood flow in mice heart imposes severe phase shift error in k-space data, resulting ghosting artifact in the image. These artifacts notably affect the accuracy of myocardial T2* relaxometry, as they introduce variable ghosting artifacts into echo images. Goal(s): To investigate the effect of Flow Compensation (FC) on myocardial T2*-map acquired by retrospective-gated Multi-Gradient-Echo (MGE) sequence. Approach: First-order GMN were added to MGE sequence. Retrospective ECG-gating was performed on murine heart. Results: FC technique can potentially reduce T2* estimation error by mitigating the flow artifact on echo images in mice heart with very fast heart rate. Impact: First-order Gradient Moment Nulling minimizes image artifacts and enhances the accuracy of myocardial T2* measurements.
Authors
- Sonia Waiczies (ORCID: https://orcid.org/0000-0002-9916-9572)
- Siqin Liu (ORCID: https://orcid.org/0000-0003-2817-892X)
- Andreas Pohlmann (ORCID: https://orcid.org/0000-0002-8572-2568)
- Min‐Chi Ku (ORCID: https://orcid.org/0000-0003-0963-2461)
- Oumaima Laghzali (ORCID: https://orcid.org/0000-0001-5923-6710)
- João Periquito (ORCID: https://orcid.org/0000-0003-3702-9264)
- Thoralf Niendorf (ORCID: https://orcid.org/0000-0001-7584-6527)
- Shahriar Shalikar (ORCID: https://orcid.org/0009-0004-1307-2166)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4551
- Citations
- 1
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.40