Rapid, accurate and precise T1 mapping in rat heart at 9.4 T MRI
Motivation: Preclinical cardiac T1 mapping in high field encountered a great challenge, such as rapid heart rate, B0 and B1 inhomogeneity, which affects both accuracy and precision in T1. Goal(s): We aim to develop an accurate, fast and reproducible T1 mapping method at 9.4 T MRI. Approach: We adapted inversion efficiency (IE) to the segmented Look-locker sequence and validated it in a phantom, as well as in aorta banding rats. Results: Segmented Look-Locker T1 mapping with IE is a fast and accurate T1 mapping method, which corrects the imperfection of inversion pulse in high field. Impact: We have developed a rapid, highly precise T1 mapping technique for rat hearts using 9.4 T MRI, enabling longitudinal scans and comparisons between different animals.
Authors
- H. Kim
- Ida Marie Hauge‐Iversen (ORCID: https://orcid.org/0000-0002-2839-1061)
- Ivar Sjaastad (ORCID: https://orcid.org/0000-0002-0131-2036)
- Einar Sjaastad Nordén
- Lili Zhang (ORCID: https://orcid.org/0000-0002-1625-7877)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3948
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00