Radial xSPEN for Non-Fourier Single-Sided MRI
Motivation: Developing methods for rapidly imaging in an inhomogeneous magnetic field is necessary for the general adoption of single sided MRI, which would bring MR imaging into the doctor's office. Goal(s): We aim to develop a fully spatiotemporal pulse sequence for collecting images with a single sided scanner. Approach: A hyperbolic phase is imparted to the magnetization which results in a time domain signal that is a profile of the phantom. The angle of this profile is rotated to produce an image. Results: Images with no Fourier encoding were produced and the images could be cropped without aliasing. Impact: This novel technique allows for rapid imaging in an inhomgeneous magnetic field. Furthermore, this technique allows for improved resolution because the field of view of the images collected with it can be smaller than the object without aliasing.
Authors
- William A. Grissom (ORCID: https://orcid.org/0000-0002-3289-1827)
- Meredith Sadinski
- Muller Gomes
- Aleksander Nacev
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2690
- Primary Topic
- Photoacoustic and Ultrasonic Imaging
- Type
- article
- Field-Weighted Citation Impact
- 0.00