Removing Background Velocity Errors in PC-MRI with Optimized Spoiler Gradient Waveforms
Motivation: Background velocity errors caused by eddy currents and mechanical oscillations in PC-MRI are a significant source of measurement error for which prospective corrections remain unavailable. Goal(s): To prospectively design spoiler gradient waveforms that reduce background velocity errors to levels generally considered irrelevant. Approach: A gradient optimization toolbox (GrOpt) was used in conjunction with a gradient impulse response to design error-minimizing waveforms. They were tested in a phantom and in 10 volunteers. Results: Background velocity errors were reduced by 84±10.4% with the gradient spoiler optimization and to levels below a clinically relevant threshold (0.4% Venc) in 96% of subjects. Impact: We used a gradient optimization (GrOpt) toolbox and the gradient impulse response function to prospectively design spoiler gradient waveforms that reduce background velocity errors in PC-MRI to levels below a clinically relevant threshold in 96% of subjects.
Authors
- Michael Loecher (ORCID: https://orcid.org/0000-0002-6533-1113)
- Daniel B. Ennis (ORCID: https://orcid.org/0000-0001-7435-1311)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4775
- Citations
- 2
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.81