In-vivo Validation of a Workflow to Predict Heating Around a Deep Brain Stimulation contacts
Motivation: To develop a reliable method to predict heating around deep brain stimulation implants as a safety assessment prior to scanning patients. Goal(s): To investigate the accuracy of a heating prediction workflow in-vivo. Approach: We surgically inserted a DBS electrode into swine brains and predicted the heating around DBS contacts, with and without perfusion. Our heating prediction workflow uses a new MR-based current measurement, proposed for this work, as well as quasi-static electromagnetic and thermal simulations to predict heating around the electrical contacts of DBS electrodes. Results: The predicted temperature around electrical contacts agreed with the measurements (NRMSE ≤ 0.09). Impact: Our workflow predicts heating around the electrical contacts in-vivo without complex modeling and simulations. The results demonstrate the reliability of the workflow to assess heating risk before scanning patients with DBS implants.
Authors
- Noam Harel (ORCID: https://orcid.org/0000-0002-6034-3801)
- Yiğitcan Eryaman (ORCID: https://orcid.org/0009-0006-7880-9872)
- Alireza Sadeghi‐Tarakameh (ORCID: https://orcid.org/0000-0001-5718-6553)
- Dee M. Koski
- Nur Izzati Huda Zulkarnain-Lemke
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3721
- Primary Topic
- Neurological disorders and treatments
- Type
- article
- Field-Weighted Citation Impact
- 0.00