Improved Detection of Target Metabolites in Brain Tumors with Intermediate TE High SNR and High Bandwidth Spin Echo Sequence at 5.0T
Motivation: challenges emerge at ultra-high fields when measuring metabolites using 1H-MRS. Goal(s): To investigate how well the high SNR and high bandwidth spin echo (HISE) technique performs at 5.0T for detecting target metabolites in brain tumors. Approach: 26 Subjects suspected of having brain tumors were enrolled. HISE and point-resolved spectroscopy (PRESS) single-voxel spectroscopy (SVS) scans were collected with a 5.0T clinical scanner with an intermediate echo time (TE=144ms). Results: HISE outperformed the clinical standard PRESS technique in detecting target metabolites of brain tumors at 5.0T, particularly Lac and Ala. Impact: In a recently developed whole-body 5.0T clinical scanner, the HISE technique has been demonstrated to be more preferable than PRESS for the clinical diagnosis of brain tumors.
Authors
- Xiaopeng Song (ORCID: https://orcid.org/0000-0003-0074-6489)
- Haibo Xu (ORCID: https://orcid.org/0000-0002-8451-8979)
- Huan Li (ORCID: https://orcid.org/0000-0001-9415-493X)
- Dan Xu (ORCID: https://orcid.org/0000-0001-9128-5328)
- Wenbo Sun (ORCID: https://orcid.org/0000-0002-7364-4419)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3858
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00