Reduction of Radiofrequency Induced Heating around Passive Implants via Flexible Metasurface Shielding at 7T

Motivation: Metallic implant compatibility at ultra-high field strengths (≥7T) continues to be often contraindicated as RF induced heating can result in surrounding tissue damage. Metasurface technology has been shown in the past to locally null the B1+ field, thereby providing a potential solution for shielding implants. Goal(s): To demonstrate a metasurface based method for shielding metallic implants to reduce RF heating. Approach: Eight implants were tested in a polyacrylic acid (PAA) phantom using a high-SAR sequence with and without a prototype metasurface. Results: On average the metasurface design reduced RF induced implant heating in the phantom by 41.6%. Impact: Patient undergoing invasive brain or trauma surgeries typically have passive metallic devices placed, making them unable to receive MRI scans at ultra-high field strengths due to RF induced heating. This work benefits this patient population by reducing RF heating.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/1215
Primary Topic
Ultrasound and Hyperthermia Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Reduction of Radiofrequency Induced Heating around Passive Implants via Flexible Metasurface Shielding at 7T

Ravinder Reddy, Wyger M. Brink, Samir Mehta, Neil Wilson et al.
1 links
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Ultrasound and Hyperthermia Applications
article

Reduction of Radiofrequency Induced Heating around Passive Implants via Flexible Metasurface Shielding at 7T

Ravinder Reddy, Wyger M. Brink, Samir Mehta, Neil Wilson, Anshuman Swain, Aniketh Hanumapur, Mark Elliott, Paul Jacobs, Neeraj Panchal
article en

Abstract

Motivation: Metallic implant compatibility at ultra-high field strengths (≥7T) continues to be often contraindicated as RF induced heating can result in surrounding tissue damage. Metasurface technology has been shown in the past to locally null the B1+ field, thereby providing a potential solution for shielding implants. Goal(s): To demonstrate a metasurface based method for shielding metallic implants to reduce RF heating. Approach: Eight implants were tested in a polyacrylic acid (PAA) phantom using a high-SAR sequence with and without a prototype metasurface. Results: On average the metasurface design reduced RF induced implant heating in the phantom by 41.6%. Impact: Patient undergoing invasive brain or trauma surgeries typically have passive metallic devices placed, making them unable to receive MRI scans at ultra-high field strengths due to RF induced heating. This work benefits this patient population by reducing RF heating.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 80%
Ultrasound and Hyperthermia Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.