A combined application of Deep Learning Recon and MRI mute technique
Motivation: MRI mute technique reduces scanning noise by lowering the gradient slew rate which increase echo spacing, resulting in image blurring and longer scan time. Goal(s): To design a scanning method that simultaneously reduces scan time and scanning noise without compromising image SNR and clarity. Approach: Develop a DL-Recon post-processing model to enhance SNR and clarity of images which are acquired from optimized knee scanning protocol with mute technique. Results: The proposed scanning method improves SNR about 38% and reduces scan time and noise sound pressure separately about 44.8% and 86%. Impact: Combining DL-Recon with mute sequences help doctors to diagnose more patients. It also enhances the success of clinical scans by decreasing scan noise to improve patient comfort. This is a successful application of DL-Recon with mute scanning technique.
Authors
- Wei Xi (ORCID: https://orcid.org/0000-0001-9348-2982)
- Weinan Tang
- Hongbin Wang (ORCID: https://orcid.org/0000-0003-2176-2998)
- Jianghua Wu (ORCID: https://orcid.org/0000-0003-3277-5639)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4660
- Primary Topic
- Neural Networks and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00