Four-Channel MM-Wave Radar Testbed for Neck vs Chest Cardiac Sensing
Motivation: Respiratory and cardiac gating are common in modern MRI, but require additional sensors in contact with the patient and the associated cabling. Goal(s): Our goal is to determine if CW mm-wave radar can act as a robust non-contact cardiac and respiratory sensor, with potential for gating in MRI. Approach: A 4-channel 24 GHz radar test bed is developed with a pair of neck radars and a pair of chest radars. SSA and ICA methods help extract waveform structure relative to an ECG. Results: The CW radars nicely separate breathing and heart rate with distinct features at the ECG QRS complex. Impact: MM-wave radar could provide a non-contact sensor of cardiac/breathing in MRI. Potential applications include non-contact gating, and independent vascular pulse or motion sensing for use in neuroimaging.
Authors
- John M. Pauly (ORCID: https://orcid.org/0000-0001-5918-4172)
- Lauren Hughes
- Fraser Robb (ORCID: https://orcid.org/0000-0002-1625-9709)
- Shreyas Vasanawala (ORCID: https://orcid.org/0000-0002-1999-6595)
- Greig Scott (ORCID: https://orcid.org/0000-0002-6609-5431)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0677
- Primary Topic
- Non-Invasive Vital Sign Monitoring
- Type
- article
- Field-Weighted Citation Impact
- 0.00