Investigation of RF-Induced Heating of Active implantable Medical Devices with Shorted Leads Under 1.5T MRI

Motivation: In active implantable medical devices (AIMDs), electrical shorts may occur either between electrodes of two leads in multi-lead systems (inter-lead shorts), or within the AIMD header due to incomplete lead insertion (intra-lead shorts). Goal(s): Whether these shorts invalidate the originally-established MR safety condition for RF-induced heating?Approach: The RF-induced heating for three shorted-lead cases were investigated. Device heating models (transfer functions) and RF-induced heating were measured for 1.5T MR systems. Results: Though the intra-lead short case exhibited a similar transfer function (TF) and RF-induced heating to non-shorted leads, inter-lead shorts had significant impact on the RF-induced heating. Impact: This research improves MRI safety for patients with active implantable medical devices (AIMDs) by studying the effects of electrical shorts.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/3740
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Investigation of RF-Induced Heating of Active implantable Medical Devices with Shorted Leads Under 1.5T MRI

Matthew Wilcox, Qingyan Wang, Ji Chen, Wenhao Zhang et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Investigation of RF-Induced Heating of Active implantable Medical Devices with Shorted Leads Under 1.5T MRI

Matthew Wilcox, Qingyan Wang, Ji Chen, Wenhao Zhang, Lijian Yang
article en

Abstract

Motivation: In active implantable medical devices (AIMDs), electrical shorts may occur either between electrodes of two leads in multi-lead systems (inter-lead shorts), or within the AIMD header due to incomplete lead insertion (intra-lead shorts). Goal(s): Whether these shorts invalidate the originally-established MR safety condition for RF-induced heating?Approach: The RF-induced heating for three shorted-lead cases were investigated. Device heating models (transfer functions) and RF-induced heating were measured for 1.5T MR systems. Results: Though the intra-lead short case exhibited a similar transfer function (TF) and RF-induced heating to non-shorted leads, inter-lead shorts had significant impact on the RF-induced heating. Impact: This research improves MRI safety for patients with active implantable medical devices (AIMDs) by studying the effects of electrical shorts.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 64%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies