Investigation of RF-Induced Heating of Active implantable Medical Devices with Shorted Leads Under 1.5T MRI
Motivation: In active implantable medical devices (AIMDs), electrical shorts may occur either between electrodes of two leads in multi-lead systems (inter-lead shorts), or within the AIMD header due to incomplete lead insertion (intra-lead shorts). Goal(s): Whether these shorts invalidate the originally-established MR safety condition for RF-induced heating?Approach: The RF-induced heating for three shorted-lead cases were investigated. Device heating models (transfer functions) and RF-induced heating were measured for 1.5T MR systems. Results: Though the intra-lead short case exhibited a similar transfer function (TF) and RF-induced heating to non-shorted leads, inter-lead shorts had significant impact on the RF-induced heating. Impact: This research improves MRI safety for patients with active implantable medical devices (AIMDs) by studying the effects of electrical shorts.
Authors
- Matthew Wilcox (ORCID: https://orcid.org/0000-0002-8070-6075)
- Qingyan Wang (ORCID: https://orcid.org/0000-0001-6104-4048)
- Ji Chen (ORCID: https://orcid.org/0000-0001-6944-2729)
- Wenhao Zhang (ORCID: https://orcid.org/0000-0002-9546-5550)
- Lijian Yang (ORCID: https://orcid.org/0000-0002-3126-2703)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3740
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00