Investigation of RF-induced Heating of Active Implantable Medical Device in the Vicinity of Fragmented Leads under 1.5T MRI
Motivation: Electrically short fragmented leads are sometimes left behind in the human body after the extraction of an active implantable medical device (AIMD). When a new AIMD is implanted near the fragmented leads, the radiofrequency (RF)-induced heating of the newly implanted AIMD could be altered. Goal(s): Develop a method to evaluate the RF-induced heating of AIMD with fragmented leads in the vicinity. Approach: We propose to use the electric field distribution which includes the presence of the fragmented leads with the AIMD transfer function model for RF-induced heating evaluation. Results: The proposed method can effectively predict the RF-induced heating for AIMDs near fragmented leads. Impact: It was observed that the RF-induced heating for AIMD system can be altered by the nearby fragmented leads.
Authors
- Ran Guo (ORCID: https://orcid.org/0000-0001-7948-4260)
- Krishna Kurpad
- Jianfeng Zheng (ORCID: https://orcid.org/0000-0001-8998-9561)
- Qingyan Wang (ORCID: https://orcid.org/0000-0001-6104-4048)
- Lijian Yang (ORCID: https://orcid.org/0000-0002-3126-2703)
- Ji Chen (ORCID: https://orcid.org/0000-0003-4401-4714)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3727
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00