Robust inter-blade and inter-slice motion correction reconstruction for PROPELLER MRI
Motivation: PROPELLER MRI has been widely used to mitigate patient motion; however, conventional approach is not very robust, and does not address inter-slice misalignment that tends to occur on few moving patients. Goal(s): Develop new robust inter-blade and inter-slice motion correction technique. Approach: We propose a novel inter-blade and inter-slice motion correction reconstruction technique for PROPELLER MRI using volumetric calibration scan as reference. Results: The results show large improvement brought by the proposed method compared to convention Motion Correction on both in-plane motion artifact mitigation and inter-slice alignment. Impact: Reducing repeat scans due to patient motion, hence, increase patient throughput. Enable scanning patient who cannot keep still in the scanner, such as pediatric patients.
Authors
- Ajeet Gaddipati
- Ali Ersöz
- Shaorong Chang
- Xucheng Zhu (ORCID: https://orcid.org/0000-0002-5537-5542)
- Min Wei (ORCID: https://orcid.org/0000-0003-3327-406X)
- Piero Ghedin
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4647
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00