Impact of Matching Network Losses on RF Coil Q and SNR at Low Field Strengths.
Motivation: RF Coil design for low frequencies. Goal(s): Demonstrate the impact of losses in different matching network topologiesApproach: we compared three- and four-element matching networks on 12 cm diameter coils at 10, 23.55, and 63.6 MHz. and evaluated coil loss and SNR. Results: At 10 and 23.55MHz, the coils with three-element (single stage) matching networks exhibited approximately 50% lower unloaded Q compared to coils with four-element matching networks, whereas unloaded Q values were within 3% for three- and four-element matched coils at 63.6MHz. SNR at 23.55MHz (0.55T) was approximately 30% higher for the four-element matched coil over the three-element matched coil. Impact: Resistive losses in four-element matching networks can be significantly lower than those of three-element networks at low frequencies or for lightly loaded coils, resulting in higher quality factor and SNR.
Authors
- Karthik Lakshmanan (ORCID: https://orcid.org/0000-0003-0187-495X)
- Ryan Brown (ORCID: https://orcid.org/0000-0002-2683-712X)
- Bili Wang (ORCID: https://orcid.org/0000-0002-9278-525X)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1403
- Primary Topic
- Microwave and Dielectric Measurement Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00