Accelerating CEST MRI using Data-Driven Z-Spectral Compressed Sensing
Motivation: CEST MRI requires the collection of multiple saturated images with different saturation offsets, resulting in prolonged scan times, which hinders its clinical applications. Goal(s): we aim to reduce the scan time of CEST MRI by recovering the undersampled Z-spectrum to full-sampling counterpart using data-driven Z-spectral compressed sensing method. Approach: The modified U-Net was employed for Z-spectral recovery. Training data were generated using Bloch equation. Numerical simulations and in vivo experiments on rat brains were conducted to validate the proposed method. Results: The results demonstrate that our method outperformed conventional interpolation methods, and threefold undersampling rate can be achieved without discernible degradation in quantification. Impact: The proposed method can efficiently reduce the scan time of CEST MRI, potentially facilitating its clinical applications.
Authors
- Tao Gong (ORCID: https://orcid.org/0000-0002-0587-7544)
- Lin Chen (ORCID: https://orcid.org/0000-0003-1397-6769)
- Haipeng Xu (ORCID: https://orcid.org/0009-0004-9277-523X)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4476
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00