Accelerating MRI with a Wireless Insert Gradient Coil
Motivation: High performance gradients enable fast and high-resolution imaging but are costly and cumbersome to install in an existing MRI-systemGoal(s): Present a wireless (resonant) gradient coil setup that yields additional gradient performance without needing a dedicated amplifierApproach: The increased gradient performance was measured using field camera measurements and acceleration performance was assessed retrospectively using phantom experimentsResults: The wireless gradient yielded a factor 23 improvement in slew rate (from 125 T/m/s to 2900 T/m/s) and 28-fold retrospective acceleration resulted in aliasing free images. Impact: A wireless insert gradient coil enables 28-fold accelerated scanning without a supplementary gradient amplifier. This provides a cost-effective pathway for improving gradient performance with minimal system modifications.
Authors
- Jeroen C.W. Siero (ORCID: https://orcid.org/0000-0001-5079-2868)
- Dennis W. J. Klomp (ORCID: https://orcid.org/0000-0002-5884-5386)
- Thomas Roos (ORCID: https://orcid.org/0000-0002-3465-7794)
- Edwin Versteeg (ORCID: https://orcid.org/0000-0003-3235-3970)
- Rik Weersink
- Sven Nouwens
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/0911
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00