Optimization of 3D PCASL perfusion SNR efficiency at 7T
Motivation: 3D-PCASL is a well-established perfusion MRI technique at 3T. Nonetheless, its effectiveness is hampered by increased RF power deposition at 7T limiting the SNR efficiency and labeling duration. Goal(s): This work investigates the relationship between gray matter perfusion signal and labeling duration aiming to optimize 3D-PCASL SNR efficiency while taking into consideration the increased RF power constraints of 7T MRI. Approach: SNR efficiency was measured at various labeling durations maintaining scan time to ~6 min and SAR to ~90% of first-level mode limit at 7T. Results: Optimal SNR efficiency was obtained at labeling duration of ~3 s. Impact: The result of this work provides optimal labeling duration for 7T 3D PCASL studies with maximum SNR per unit time taking into account first-level mode RF power limits.
Authors
- Alan P. Koretsky (ORCID: https://orcid.org/0000-0002-8085-4756)
- Gaël Saïb (ORCID: https://orcid.org/0000-0003-2571-2410)
- S. Lalith Talagala
- Sunil Patil
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2003
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00