Reduction of RF-heating on bilateral DBS leads using two channel RF-shimming on 3T MRI
Motivation: Bilateral deep brain stimulation (DBS) surgery would benefit from safe intraoperative 3T MRI using conventional scanners. Goal(s): Our goal was to develop a method to determine patient-specific safety constraints to scan a bilateral lead configuration with minimal RF-heating, by minimizing RF-induced currents at maximal B1. Approach: We scanned a phantom with two leads in different configurations, measured local heating and determined the transimpedance. Results: We could determine the phantom-specific safety constraints and find a low-SAR sequence meeting the constraints, after optimizing the two channel RF-shimming to minimize RF-heating by minimizing RF-induced currents at maximal B1. Impact: The proposed method can be used to minimize heating of a bilateral lead configuration and estimate patient-specific safety constraints, but further research has to be conducted to be able to use it during deep brain stimulation surgery.
Authors
- Wyger M. Brink (ORCID: https://orcid.org/0000-0001-9974-7662)
- Ruud Vinke
- Marinette van der Graaf (ORCID: https://orcid.org/0000-0001-7495-987X)
- C.D.E. Van Speybroeck
- W. Roskamp
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/3726
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00