Geometry Matters: High Performance Acceleration with Twisted Pair and Conventional Coil Designs at 7T MRI
Motivation: Driven by the potential of the highly flexible twisted pair coil in creating unique sensitivity patterns for optimal acceleration. Goal(s): To compare geometry factors of the twisted pair and conventional copper coils, exploring effective arrangements for various coil shapes. Approach: Simulations of eight-element arrays containing circular and elongated coils; assessing geometry factors and the impact of random coil placements. Results: Findings emphasize similar geometry factors between the twisted pair and conventional coils. Moreover, random coil setups around a phantom do not inherently improve geometry factors. Impact: A comparison between twisted pair and conventional coils shows similar performance in g-factors. The flexibility of the twisted pair can be used to improve the g-factor but more work is required in finding the optimal shape of these coils.
Authors
- Irena Zivkovic (ORCID: https://orcid.org/0000-0001-6900-6580)
- Jules Vliem (ORCID: https://orcid.org/0009-0008-6680-6345)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1439
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00