Optimisation and implementation of high b-value DWI using 200mT/m gradient system for clinical prostate MRI.

Motivation: Diffusion-weighted imaging is critical to the diagnosis of prostate cancer but can be limited by noise. A commercially available 200mT/m gradient system may allow revisiting of higher b-value DWI. Goal(s): To optimise a feasible high b-value prostate MR protocol for clinical use. Approach: Phantom, simulation and in vivo iterative optimisation were evaluated against technical metrics and novel discrete choice experimental design for radiologist preference. Results: Radiologist DWI preference is weighted by signal-to-noise. Emerging high gradient performance and deep learning reconstruction can reduced decline in SNR potentiating a signal-optimised high b-value protocol for clinical prostate cancer imaging within feasible scan times. Impact: Signal-optimised high b-value DWI for prostate cancer using a 200mT/m gradient 3T MRI system may be clinically feasible and supports a scalable trial of its diagnostic impact.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4321
Citations
1
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.40
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Optimisation and implementation of high b-value DWI using 200mT/m gradient system for clinical prostate MRI.

Mitchell C. Raeside, Andrew Dwyer, Kirsten Gormly, Stephanie Withey et al.
1 citations
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
0.40
article

Optimisation and implementation of high b-value DWI using 200mT/m gradient system for clinical prostate MRI.

Mitchell C. Raeside, Andrew Dwyer, Kirsten Gormly, Stephanie Withey, Angela Walls
article en
1 citations

Abstract

Motivation: Diffusion-weighted imaging is critical to the diagnosis of prostate cancer but can be limited by noise. A commercially available 200mT/m gradient system may allow revisiting of higher b-value DWI. Goal(s): To optimise a feasible high b-value prostate MR protocol for clinical use. Approach: Phantom, simulation and in vivo iterative optimisation were evaluated against technical metrics and novel discrete choice experimental design for radiologist preference. Results: Radiologist DWI preference is weighted by signal-to-noise. Emerging high gradient performance and deep learning reconstruction can reduced decline in SNR potentiating a signal-optimised high b-value protocol for clinical prostate cancer imaging within feasible scan times. Impact: Signal-optimised high b-value DWI for prostate cancer using a 200mT/m gradient 3T MRI system may be clinically feasible and supports a scalable trial of its diagnostic impact.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 32%
Advanced MRI Techniques and Applications
0.40
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.