Fast and High-Resolution Intracranial MR Angiography at 7T using FLEXA
Motivation: Non-contrast 3DTOF demonstrates visualizing the cerebral microvasculature at 7T MRI, but the scan time is too long, typically about 10 minutes due to requiring high spatial resolution.Goal(s): Our goal was to reduce the acquisition time by a factor of 2.Approach: We utilize FLEXA on the high gradient performance system at 7T.Results: Comparable image quality to 3DTOF was obtained with FLEXA technique with a factor of 2 decrease in acquisition time.Impact: The improvement in acquisition speed through FLEXA comes from the strategy of short TR with thin slab and opens new standard for scan protocol optimization at 7T MRA.
Authors
- Naoyuki Takei (ORCID: https://orcid.org/0000-0003-1171-8256)
- Nathan James Roberts (ORCID: https://orcid.org/0000-0001-8299-4499)
- Xucheng Zhu (ORCID: https://orcid.org/0000-0002-5537-5542)
- Tetsuya Wakayama
- Brian Burns (ORCID: https://orcid.org/0009-0002-0318-8400)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4885
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00