Preliminary study of a receive Multi-Loop Coil (MLC) array for 7T brain MRI
Motivation: The use of MLCs has shown the potential to improve the SNR at short distance as compared to an equivalent SLC. Goal(s): Evaluation of the performances of a 8-channels MLC-array for head imaging at 7T. Approach: Electromagnetic simulation was used to evaluate and compare the SNR, noise-covariance matrix and g-maps obtained with the MLC-array and with an equivalent array of SLC. Results: As compared to the SLC-array, the MLC-array achieves an increased SNR in a relatively large peripheral ring and a reduced maximum g-factor. Impact: Array of MLC represents a valuable strategy for array developpement at high field that can be employed to improve the SNR or reduce the number of channels.
Authors
- Alexis Amadon
- Marie Poirier‐Quinot (ORCID: https://orcid.org/0000-0003-3199-7432)
- Paul‐François Gapais (ORCID: https://orcid.org/0000-0002-6592-5034)
- Jean‐Christophe Ginéfri (ORCID: https://orcid.org/0000-0002-3545-995X)
- Alexandre Vignaud
- Clément Thibault
- C. Dubuc
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1445
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00