Assessing RF-induced heating safety of MRI scans with shoulder implants at 7T

Motivation: Brain MRI with shoulder implants are often contraindicated at 7T with head coils due to possibility of radiofrequency (RF)-induced heating. Such implants are not uncommon in older adults; contraindication results in increasing proportion of elderly to be denied high-field scans. Assessing heating safety of MRI at 7T with implants is of significance. Goal(s): To measure shoulder implant heating during MRI at 7T with pTx head coils and assess safety thereof. Approach: Temperature of a reverse total shoulder implant was measured with fluoroptic sesnors during MRI at Siemens 7T Terra with pTX head coil. Results: No significant heating (>1°C) of the implant was observed. Impact: Findings from this study may enable brain scanning individuals with shoulder implant at 7T, thus making improved image quality a possibility.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/1448
Primary Topic
Advanced MRI Techniques and Applications
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Assessing RF-induced heating safety of MRI scans with shoulder implants at 7T

Jon M Dickson, Pallab Bhattacharyya, Mark E. Lowe, Stephen E. Jones et al.
1 links
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced MRI Techniques and Applications
article

Assessing RF-induced heating safety of MRI scans with shoulder implants at 7T

Jon M Dickson, Pallab Bhattacharyya, Mark E. Lowe, Stephen E. Jones, Jason Ho
article en

Abstract

Motivation: Brain MRI with shoulder implants are often contraindicated at 7T with head coils due to possibility of radiofrequency (RF)-induced heating. Such implants are not uncommon in older adults; contraindication results in increasing proportion of elderly to be denied high-field scans. Assessing heating safety of MRI at 7T with implants is of significance. Goal(s): To measure shoulder implant heating during MRI at 7T with pTx head coils and assess safety thereof. Approach: Temperature of a reverse total shoulder implant was measured with fluoroptic sesnors during MRI at Siemens 7T Terra with pTX head coil. Results: No significant heating (>1°C) of the implant was observed. Impact: Findings from this study may enable brain scanning individuals with shoulder implant at 7T, thus making improved image quality a possibility.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Good health and well-being
Openalex Percentile: Top 64%
Advanced MRI Techniques and Applications
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.