B1 SHIM CALIBRATION USING 1D ENCODED B1 DATA

Motivation: To develop a method for fast per-patient calibration of B1 shim parameters. Goal(s): To reduce scan time for per-patient B1 shimming calibration to improve patient throughput. Approach: A key insight is that 1D projections acquired with specific projection orientations are sufficient to capture B1 distribution patterns. By comparing a pair of orthogonal B1 projections in the brain or by measuring average B1 amplitude of a pair of ROIs in a single projection across the breast, left-right B1 symmetry can be measured. Results: Optimal B1 shim settings can be determined with a total time < 1 second as confirmed by B1 maps. Impact: The proposed method for per-patient B1 shimming calibration using 1D B1 projections instead of 2D B1 maps can quickly calibrate B1 shimming parameters in less than one second.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4946
Primary Topic
Medical Image Segmentation Techniques
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

B1 SHIM CALIBRATION USING 1D ENCODED B1 DATA

W Dannels, Andrew Wheaton
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Medical Image Segmentation Techniques
article

B1 SHIM CALIBRATION USING 1D ENCODED B1 DATA

W Dannels, Andrew Wheaton
article en

Abstract

Motivation: To develop a method for fast per-patient calibration of B1 shim parameters. Goal(s): To reduce scan time for per-patient B1 shimming calibration to improve patient throughput. Approach: A key insight is that 1D projections acquired with specific projection orientations are sufficient to capture B1 distribution patterns. By comparing a pair of orthogonal B1 projections in the brain or by measuring average B1 amplitude of a pair of ROIs in a single projection across the breast, left-right B1 symmetry can be measured. Results: Optimal B1 shim settings can be determined with a total time < 1 second as confirmed by B1 maps. Impact: The proposed method for per-patient B1 shimming calibration using 1D B1 projections instead of 2D B1 maps can quickly calibrate B1 shimming parameters in less than one second.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 79%
Medical Image Segmentation Techniques
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.