A Multi-Directional 4D PCMR Flow Phantom: Design and Experiments
Motivation: Accurate blood flow measurement is vital for the diagnosis of cardiovascular disorders. Existing phantoms focus primarily on validating flow using phase-contrast MR (PCMR) techniques along the three orthogonal directions. No phantom has been reported for the validation of flow in non-orthogonal orientations. Goal(s): To validate flow quantification accuracy on a phantom in non-orthogonal directions using 4D flow MRI. Approach: A multi-directional plexiglass phantom was designed and fabricated to evaluate accuracy of 4D flow PCMR using blood-mimicking flows within tubing controlled by a programmable pump. Results: The phantom was successfully used to validate flow in orthogonal and non-orthogonal directions simultaneously with 4D flow MRI. Impact: The flow phantom can be used to validate flow in 5 directions simultaneously with 4D flow MRI. The phantom has the potential to be used to optimize and standardize the 4D flow MR protocol parameters for clinical applications.
Authors
- Enamul Bhuiyan (ORCID: https://orcid.org/0000-0002-3098-1095)
- Guangyu Dan (ORCID: https://orcid.org/0000-0003-0427-5661)
- Lauren Ostergren
- Xiaohong Joe Zhou (ORCID: https://orcid.org/0000-0003-0793-4925)
- Muge Karaman (ORCID: https://orcid.org/0000-0002-7080-7307)
- Fady T. Charbel
- Kezhou Wang (ORCID: https://orcid.org/0000-0003-1006-9522)
- Qingfei Luo (ORCID: https://orcid.org/0000-0001-5574-817X)
- Taylor Otey
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4771
- Primary Topic
- Aerodynamics and Acoustics in Jet Flows
- Type
- article
- Field-Weighted Citation Impact
- 0.00