Value of placental stiffness using virtual magnetic resonance elastography in pregnancies complicated by pre-eclampsia

Motivation: To evaluate virtual magnetic resonance elastography in healthy and preeclamptic(PE) pregnancies. Goal(s): To compare the stiffness value (μdiff) and apparent diffusion coeffificient (ADC) in healthy and PE pregnancies. Approach: DWI(b-value of 50 ,200and 800 s/mm2 ) were performed on all pregnant women using a 1.5 T MRI scanner. The value of μdiff and ADC were calculated and compared between groups. Results: The mean ADC value of the control and PE groups were 1.414±0.228×10-3 mm2/s and 1.737±0.107×10-3 mm2/s, respectively. The mean μdiff value of PE were 5.901±1.757 and 4.618±2.055 kPa, respectively. The area under the curve for μdiff was 0.903 and ADC was 0.796, respectively. Impact: Placental μdiff value was found to be more reliable than ADC in differentiating between normal and preeclampsia placentas.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/4819
Primary Topic
Ultrasound Imaging and Elastography
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Value of placental stiffness using virtual magnetic resonance elastography in pregnancies complicated by pre-eclampsia

Jiejun Cheng, Jialu Xu
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Ultrasound Imaging and Elastography
article

Value of placental stiffness using virtual magnetic resonance elastography in pregnancies complicated by pre-eclampsia

Jiejun Cheng, Jialu Xu
article en

Abstract

Motivation: To evaluate virtual magnetic resonance elastography in healthy and preeclamptic(PE) pregnancies. Goal(s): To compare the stiffness value (μdiff) and apparent diffusion coeffificient (ADC) in healthy and PE pregnancies. Approach: DWI(b-value of 50 ,200and 800 s/mm2 ) were performed on all pregnant women using a 1.5 T MRI scanner. The value of μdiff and ADC were calculated and compared between groups. Results: The mean ADC value of the control and PE groups were 1.414±0.228×10-3 mm2/s and 1.737±0.107×10-3 mm2/s, respectively. The mean μdiff value of PE were 5.901±1.757 and 4.618±2.055 kPa, respectively. The area under the curve for μdiff was 0.903 and ADC was 0.796, respectively. Impact: Placental μdiff value was found to be more reliable than ADC in differentiating between normal and preeclampsia placentas.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Good health and well-being
Openalex Percentile: Top 64%
Ultrasound Imaging and Elastography
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.

Value of placental stiffness using virtual magnetic resonance elastography in pregnancies complicated by pre-eclampsia — Jiejun Cheng, Jialu Xu · Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition (2024) | TGRS Research Map | TGRS