A Novel Transceive 2-channel Orthogonally Arranged Dual Loop Solenoid RF Coil Design for Improving Low Field MRI SNR
Motivation: There are growing interest in low field specialised MRI R&D but they are currently burdened by their low SNR images. Goal(s): A novel 2-channel orthogonally arranged dual loop solenoid RF coil design has been proposed to improve low field MR image SNR. Approach: Two transceive 2-Channel RF coil prototypes using the novel design and using conventional design were constructed and tested in a 0.35T MRI system. Results: In comparing the acquired MR images, it is shown that the proposed novel design can improve low field MR image SNR by ≥20% as compared to the conventional design. Impact: A novel 2-channel orthogonally arranged dual loop solenoid transceive RF coil has been constructed and tested in a 0.35T MRI system. Using this new RF coil design it is shown that the acquired image SNR can be increased by ≥20%.
Authors
- Bing Keong Li
- Jian Bao
- Shao Ying Huang (ORCID: https://orcid.org/0000-0003-3775-8205)
- Y. J. Gu (ORCID: https://orcid.org/0000-0002-6234-8489)
- Xin Qi (ORCID: https://orcid.org/0000-0002-5874-7321)
- H. Pei (ORCID: https://orcid.org/0000-0002-5078-3336)
- Ren Jie Zong
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1599
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00