Myocardial perfusion measurements with a deep learning-assisted cardiac arterial spin labeling (DeepCASL): towards validation by microsphere
Motivation: Cardiac arterial spin labeling (ASL) method is sensitive to noise (system and physiology), which may lead to inaccurate MBF measurement. Goal(s): A cardiac MRI arterial spin labeling method was developed with assistance of a deep learning networks (DeepCASL) to improve image quality and measurement accuracy. Approach: The performance of the DeepCASL method was evaluated in a canine model of coronary arterial disease by comparing and correlating with MBF determined by microsphere measurements. Results: The validation study revealed moderate to strong correlations in absolute myocardial blood flow values between MRI and microsphere reference methods. Impact: This new DeepCASL technique opens a door for clinical applications of noncontrast cardiac perfusion as a screen tool for reliable diagnosis of perfusion deficit in a variety of cardiomyopathy disorders.
Authors
- Ran Li (ORCID: https://orcid.org/0000-0002-9863-4110)
- Jie Zheng (ORCID: https://orcid.org/0000-0002-9640-1604)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/1510
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00