Bias and Variability of Mean Diffusivity from DTI Across Two Head Coils: a Traceable Phantom Study
Motivation: Variability and bias of diffusion parameters need to be assessed to separate the measurement-induced variability from biological effects. Goal(s): Our goal was to assess the bias and variability of mean diffusivity (MD) from the Diffusion Tensor Model using a traceable phantom. Approach: The NIST/QIBA phantom was scanned multiple times using two different head coils on the same system. The differences between MD measured using different head coils and bias to NIST-reported values were investigated. Results: A significant difference in measured MD with two coils can be found across multiple vials. Both coils introduce non-negligible bias to NIST-reported values for lower MD values. Impact: Using traceable phantoms is essential in the development of Diffusion MRI-based Quantitative Imaging Biomarkers (QIBs). Different head coils can introduce significant differences across MD values and should be treated as confounding factors in QIBs studies.
Authors
- Agnieszka Sierhej
- Matt G. Hall (ORCID: https://orcid.org/0000-0002-9530-5477)
- Nadia Smith (ORCID: https://orcid.org/0000-0002-6503-3001)
- Chris Clark
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4893
- Primary Topic
- Thermography and Photoacoustic Techniques
- Type
- article
- Field-Weighted Citation Impact
- 0.00