Initial results in pediatric volunteers using a silent and motion corrected ZTE protocol

Motivation: Improve imaging for pediatric subjects by reducing loud acoustic noise and accommodating subject motion. Goal(s): Develop a ZTE-based silent and motion-corrected neuroimaging protocol in pediatric subjects. Approach: Nine pediatric volunteers were imaged with a protocol consisting of silent ZTE based T1w and T2w acquisitions and conventional Cartesian 3D gradient echo T1 MPRAGE and T2 FSE-CUBE. Images were reviewed and scored by a board-certified radiologist to assess overall image quality. Results: Radiologist review indicates relatively comparable image quality for the ZTE acquisitions versus the conventional acquisitions. Motion induced blurring of some brain structures during motion was reduced using ZTE with motion correction. Impact: This approach shows promise for studies in sensitive populations such as pediatrics with ear disorders and autism spectrum disorder.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/2535
Primary Topic
Hemodynamic Monitoring and Therapy
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Initial results in pediatric volunteers using a silent and motion corrected ZTE protocol

James H. Holmes, Mathews Jacob, Vincent A. Magnotta, Curtis A. Corum et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Hemodynamic Monitoring and Therapy
article

Initial results in pediatric volunteers using a silent and motion corrected ZTE protocol

James H. Holmes, Mathews Jacob, Vincent A. Magnotta, Curtis A. Corum, Assim Saad Eddin, Yan Chen, J.E. Hanson, Paul Dicamillo
article en

Abstract

Motivation: Improve imaging for pediatric subjects by reducing loud acoustic noise and accommodating subject motion. Goal(s): Develop a ZTE-based silent and motion-corrected neuroimaging protocol in pediatric subjects. Approach: Nine pediatric volunteers were imaged with a protocol consisting of silent ZTE based T1w and T2w acquisitions and conventional Cartesian 3D gradient echo T1 MPRAGE and T2 FSE-CUBE. Images were reviewed and scored by a board-certified radiologist to assess overall image quality. Results: Radiologist review indicates relatively comparable image quality for the ZTE acquisitions versus the conventional acquisitions. Motion induced blurring of some brain structures during motion was reduced using ZTE with motion correction. Impact: This approach shows promise for studies in sensitive populations such as pediatrics with ear disorders and autism spectrum disorder.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 60%
Hemodynamic Monitoring and Therapy
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.