Renal multiple post label times arterial spin label (ASL) read out by turbo spin echo (TSE)

Motivation: Renal imaging using echo planar imaging (EPI) readout may be sensitive to field inhomogeneities and induced ghost artifact. Besides, single inversion time is not enough for renal blood flow (RBF) quantification. Investigation of more stable clinical arterial spin label (ASL) imaging methods is required. Goal(s): We aimed to design a stable clinical ASL sequence to quantify RBF accurately. Approach: Multiple inversion time (TIs) readout by turbo spin echo (TSE) was processed, and the repeatability of quantification was evaluated in volunteers (n=7). Results: Multiple TIs combined with TSE readout of FAIR-ASL provided more accurate RBF measurement and highly repeatability compare with pcASL-EPI. Impact: Multi-PLDs readout by TSE may provide a stable clinical ASL-FAIR sequences to accurately quantify RBF.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/2002
Primary Topic
Cardiac Arrhythmias and Treatments
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Renal multiple post label times arterial spin label (ASL) read out by turbo spin echo (TSE)

Zelong Chen, Zhigang Wu, Peng Wu, Kangfa Deng et al.
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Cardiac Arrhythmias and Treatments
article

Renal multiple post label times arterial spin label (ASL) read out by turbo spin echo (TSE)

Zelong Chen, Zhigang Wu, Peng Wu, Kangfa Deng, Wenyan Zhang, Qianqian Zhang, Yikai Xu, Yizhe Zhang, Quan Tao, Yanqiu Feng
article en

Abstract

Motivation: Renal imaging using echo planar imaging (EPI) readout may be sensitive to field inhomogeneities and induced ghost artifact. Besides, single inversion time is not enough for renal blood flow (RBF) quantification. Investigation of more stable clinical arterial spin label (ASL) imaging methods is required. Goal(s): We aimed to design a stable clinical ASL sequence to quantify RBF accurately. Approach: Multiple inversion time (TIs) readout by turbo spin echo (TSE) was processed, and the repeatability of quantification was evaluated in volunteers (n=7). Results: Multiple TIs combined with TSE readout of FAIR-ASL provided more accurate RBF measurement and highly repeatability compare with pcASL-EPI. Impact: Multi-PLDs readout by TSE may provide a stable clinical ASL-FAIR sequences to accurately quantify RBF.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 60%
Cardiac Arrhythmias and Treatments
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.