Motion robust MR-ARFI using timeseries single-shot spiral acquisition
Motivation: 2DFT MR-ARFI methods used to map ultrasound beam profiles can fail with small amounts of head motion because the signal is based on phase shifts from um-sized Acoustic Radiation Force displacement of tissue. Goal(s): To demonstrate that a single-shot timeseries method is more robust to bulk motion. Approach: Employing an efficient spiral redout trajectory, we compared the two methods’ ARFI images in a tissue-mimicking phantom undergoing controlled random translations. Results: The timeseries method demonstrated good quality tissue translation maps, while the 2DFT method failed with motions as small as ~50 microns RMS. Impact: Our single-shot method may allow routine use of MR-ARFI in humans for guiding US neuromodulation and US-based ablative therapy.
Authors
- Gary H. Glover (ORCID: https://orcid.org/0000-0002-0562-0474)
- Kim Butts Pauly (ORCID: https://orcid.org/0000-0002-6281-2771)
- Morteza Mohammadjavadi
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4577
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00