Design of a nonlinear human-breast gradient coil with controllable gradient variation
Motivation: To assess diffusion metrics on a cellular scale necessitates much higher gradient amplitudes. A local nonlinear breast gradient coil needs methods of controlling the field variation during the coil design stage. Goal(s): Control the y-axis gradient variation of a double layer local breast gradient coil to achieve far above 1T/m in the whole target region. Approach: Incorporation of a constraint to control the variation of the gradient along the y-axis with a double-layer current-carrying surface. Results: A non-linear local human breast gradient coil with a gradient strength between 1.61 and 3.74 T/m for a current of 650 A was designed. Impact: The coil design with a subsequent implementation paves the way to the application of novel diffusion imaging techniques for the detection and characterization of breast cancer.
Authors
- Philipp Amrein (ORCID: https://orcid.org/0000-0002-0215-1510)
- Sebastian Littin (ORCID: https://orcid.org/0000-0002-8190-1761)
- Maxim Zaitsev (ORCID: https://orcid.org/0000-0001-7530-1228)
- Feng Jia (ORCID: https://orcid.org/0000-0003-3593-2513)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/4890
- Primary Topic
- Wireless Power Transfer Systems
- Type
- article
- Field-Weighted Citation Impact
- 0.00