T2* Relaxation Time Measurements of Various Brain Regions at 0.5 T
Motivation: Accurate knowledge of T2* relaxation times is important for optimizing contrast and SNR in multi-echo gradient echo sequences. Goal(s): To measure the nominal T2* relaxation parameter at 0.5 T over subregions in the brain. Approach: A 2 mm isotropic 3D multi-echo GRE sequence was acquired in 5 healthy volunteers. T2* estimates were computed over regions of interest using a mono-exponential fit. Results: T2* estimates of gray (86±8 ms) and white (78±5 ms) matter are in good agreement with previously published measurements in the mid-field. In the deep brain, average T2* values ranged between 70±2 ms to 89±5 ms. Impact: For optimal T2* contrast in the brain at 0.5 T, echoes from multi-echo acquisitions should extend beyond 60 ms. For optimal SNR, echoes should be combined with a target T2* of ~80 ms.
Authors
- Blaine A. Chronik
- Curtis N. Wiens
- Andrea Soddu (ORCID: https://orcid.org/0000-0001-6833-745X)
- Arjama Halder (ORCID: https://orcid.org/0000-0002-3195-9623)
- Chad Harris
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2683
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00