Imaging Near Metallic Implants: Initial 0.55T vs. 3T MRI Comparison in Hip Implant Patients

Motivation: Patients with orthopedic metallic implants often require diagnostic imaging to evaluate adjacent tissues. MRI performance, including artifact levels and SNR, varies with field strength. Goal(s): To compare 0.55T and 3T MRI for imaging patients with total hip arthroplasty (THA). Approach: Patients with THA were scanned with similar protocols at 0.55T and 3T, including multi-spectral imaging (MSI). We qualitatively compared the images from both scanners. Results: Metal artifact severity was reduced at 0.55T compared to 3T at the expense of SNR. Diagnostic imaging of patients with titanium hip implants at 0.55T is possible without MSI. Impact: We qualitatively compared 0.55T and 3T image quality in patients with hip replacements. Our findings indicate that 0.55T MRI offers substantially reduced metal artifacts and advanced multi-spectral techniques may not be required in many cases.

Authors

Publication Details

Journal
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Published
2024-11-26
DOI
https://doi.org/10.58530/2024/2650
Primary Topic
Advanced X-ray and CT Imaging
Type
article
Field-Weighted Citation Impact
0.00
Controls
|||
ALL TIME
JAN
FEB
MAR
APR
MAY
JUN
JUL
AUG
SEP
OCT
article

Imaging Near Metallic Implants: Initial 0.55T vs. 3T MRI Comparison in Hip Implant Patients

Kübra Keskin, Sophia Cui, Krishna S. Nayak, Jay R. Lieberman et al.
2 links
Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Advanced X-ray and CT Imaging
article

Imaging Near Metallic Implants: Initial 0.55T vs. 3T MRI Comparison in Hip Implant Patients

Kübra Keskin, Sophia Cui, Krishna S. Nayak, Jay R. Lieberman, Zorica Buser, Bochao Li, Jordan S. Gross, Brian Hargreaves
article en

Abstract

Motivation: Patients with orthopedic metallic implants often require diagnostic imaging to evaluate adjacent tissues. MRI performance, including artifact levels and SNR, varies with field strength. Goal(s): To compare 0.55T and 3T MRI for imaging patients with total hip arthroplasty (THA). Approach: Patients with THA were scanned with similar protocols at 0.55T and 3T, including multi-spectral imaging (MSI). We qualitatively compared the images from both scanners. Results: Metal artifact severity was reduced at 0.55T compared to 3T at the expense of SNR. Diagnostic imaging of patients with titanium hip implants at 0.55T is possible without MSI. Impact: We qualitatively compared 0.55T and 3T image quality in patients with hip replacements. Our findings indicate that 0.55T MRI offers substantially reduced metal artifacts and advanced multi-spectral techniques may not be required in many cases.

Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
Openalex Percentile: Top 80%
Advanced X-ray and CT Imaging
AI Navigator

Ask Laika to Summarize, Analyze, and Connect papers live on the map.

Summarize Papers & Methodologies

Extract key findings, datasets, and comparative methods across publications.

Benchmark Rankings & Visual Analytics

Rank top research institutions, authors, funders, topics, and journals by Field-Weighted Citation Impact (FWCI) and paper volume with instant charts.

Connect Distant Disciplines

Bridge topological clusters on the map to find hidden collaborative intersections.