MRI-guided histotripsy targeting using MR-thermometry and MR-ARFI
Motivation: Transcranial histotripsy requires pre-therapy targeting to ensure treatment at the intended region of interest. MR-ARFI and MR-Thermometry are used for targeting HIFU treatments, but are not tested for targeting of transcranial histotripsy treatments since they use different ultrasound treatment parameters. Goal(s): We evaluate the accuracy of MR-thermometry and MR-ARFI for targeting histotripsy treatments in ex-vivo tissues. Approach: Low intensity MR-thermometry and MR-ARFI were done prior to histotripsy treatment on bovine brain tissues. The lesion location was compared with the peak temperature/displacement to quantify targeting error. Results: Both MR-thermometry and MR-ARFI can perform accurate targeting of histotripsy treatments with mean errors of 2mm. Impact: Transcranial histotripsy is a promising non-invasive treatment method that requires pre-treatment targeting to ensure treatment delivery at the desired location. We showed that both low temperature MR-thermometry and MR-ARFI can be used to accurately estimate histotripsy treatment location.
Authors
- Steven P. Allen (ORCID: https://orcid.org/0000-0003-4778-8168)
- Zhen Xu (ORCID: https://orcid.org/0000-0003-0262-3044)
- Timothy J. Hall (ORCID: https://orcid.org/0000-0001-9850-6031)
- Tarana Parvez Kaovasia (ORCID: https://orcid.org/0000-0002-1786-8215)
- Douglas C. Noll (ORCID: https://orcid.org/0000-0002-0983-3805)
- Dinank Gupta (ORCID: https://orcid.org/0000-0001-7496-1056)
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2709
- Primary Topic
- Spectroscopy Techniques in Biomedical and Chemical Research
- Type
- article
- Field-Weighted Citation Impact
- 0.00