Fast Single-Point Macromolecular Proton Fraction Mapping using a 0.3 T MRI System
Motivation: Quantifying myelin content in 0.3 T low-field MRI presents a challenge owing to prolonged imaging times and a low signal-to-noise ratio. Goal(s): The study aimed to validate the practical feasibility of macromolecular proton fraction (MPF) mapping in the brain using a 0.3 T MRI system. Approach: A phantom study with protein samples and an in vivo investigation were conducted. Results: The study findings indicate a robust correlation between 0.3 T MPF in brain tissue, MPF at 3.0 T, and previously reported MPF at 0.5 T. Impact: The fast single-point MPF mapping using a 0.3 T MRI exhibited the capacity to accurately measure brain MPF in a clinically feasible timeframe, providing a valuable tool for assessing myelin content.
Authors
- Hiroyuki Kumazoe (ORCID: https://orcid.org/0000-0001-5177-790X)
- Shoma Eitoku (ORCID: https://orcid.org/0000-0002-6703-1368)
- Mika Kitajima (ORCID: https://orcid.org/0000-0002-6869-3638)
- Takahisa Izumi
- Yasuhiro Fujiwara (ORCID: https://orcid.org/0000-0003-4831-444X)
- Nobutaka Sakae
- Yuuki Motoyama
Publication Details
- Journal
- Proceedings on CD-ROM - International Society for Magnetic Resonance in Medicine. Scientific Meeting and Exhibition/Proceedings of the International Society for Magnetic Resonance in Medicine, Scientific Meeting and Exhibition
- Published
- 2024-11-26
- DOI
- https://doi.org/10.58530/2024/2515
- Primary Topic
- Advanced MRI Techniques and Applications
- Type
- article
- Field-Weighted Citation Impact
- 0.00